Loading...
om 5184516-100 Rev. 5 5-15
2-2Fluoro Acquisition Screen
ILLUSTRATION 5-15
FLUORO ACQUISITION SCREEN
Selections
This screen will display record factors as well as fluoro factors.
The top portion of the screen is dedicated for record factors.
You may change your contrast type for the trajectories to compensate techni-
cal parameters. However it is recommended to use the appropriate contrast
per procedure.
The other option is frames per second. You may select single, 1, 2, 3, 4, 5, 6
or 7.5 maximum.
The lower portion of the screen is dedicated to fluoro factors.
You may select the image intensifier size. (This is system dependent.)
The Field of View options are 16, 12, 9, 6 or 12, 9, 6, 4.5 inch (40, 32, 22,
16 cm or 32, 22, 12, 9 cm) modes.
Pediatric Dose Selection (optional) - Provides specialized image acquisition
parameters for small pediatric patients. This special purchase option provides
dedicated and tailored dose trajectories for infants and small children.
The dose can be limited to one of three rates: the standard 10 R/minute grid
in or 5R/minute input dose (grid in or out).
Provides the optimum mix between minimal dose and maximum image qual-
ity.
The pediatric mode is not available in Germany or New Zealand due to exist-
ing low dose regulations.
Pulsed Fluoro (option) - The system works with 15, 7.5 and 3.75 frames per sec-
ond. After selecting, the rates can be changed at the IDD.
FOR TRAINING PURPOSES ONLY!