om 5184516-100 Rev. 5 5-15
2-2Fluoro Acquisition Screen
ILLUSTRATION 5-15
FLUORO ACQUISITION SCREEN
Selections
• This screen will display record factors as well as fluoro factors.
• The top portion of the screen is dedicated for record factors.
• You may change your contrast type for the trajectories to compensate techni-
cal parameters. However it is recommended to use the appropriate contrast
per procedure.
• The other option is frames per second. You may select single, 1, 2, 3, 4, 5, 6
or 7.5 maximum.
• The lower portion of the screen is dedicated to fluoro factors.
• You may select the image intensifier size. (This is system dependent.)
• The Field of View options are 16, 12, 9, 6 or 12, 9, 6, 4.5 inch (40, 32, 22,
16 cm or 32, 22, 12, 9 cm) modes.
Pediatric Dose Selection (optional) - Provides specialized image acquisition
parameters for small pediatric patients. This special purchase option provides
dedicated and tailored dose trajectories for infants and small children.
• The dose can be limited to one of three rates: the standard 10 R/minute grid
in or 5R/minute input dose (grid in or out).
• Provides the optimum mix between minimal dose and maximum image qual-
ity.
• The pediatric mode is not available in Germany or New Zealand due to exist-
ing low dose regulations.
Pulsed Fluoro (option) - The system works with 15, 7.5 and 3.75 frames per sec-
ond. After selecting, the rates can be changed at the IDD.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).