om 5184516-100 Rev. 5 A-20
TABLE A-2
DUTY CYCLE CHART
Exposure Time FILM FRAME RATE
Seconds 2/sec. 4/sec. 6/sec. 8/sec.
.005
.0064
.008
.010
.0125
.016
.020
.025
.032
.040
.050
.064
.080
.10
.125
.16
1.0%
1.28%
1.6%
2.0%
2.5%
3.2%
4.0%
5.0%
6.4%
8.0%
10.0%
12.8%
16.0%
20.0%
25.0%
32.0%
2.0%
2.56%
3.2%
4.0%
5.0%
6.4%
8.0%
10.0%
12.8%
16.0%
20.0%
25.6%
32.0%
40.0%
50.0%
64.0%
3.0%
3.84%
4.8%
6.0%
7.5%
9.6%
12.0%
15.0%
19.2%
24.0%
30.0%
38.4%
48.0%
60.0%
4.0%
5.12%
6.4%
8.0%
10.0%
12.8%
16.0%
20.0%
25.6%
32.0%
40.0%
51.2%
MINIMUM COOLING INTERVAL BETWEEN SERIES
kVp x mA x DUTY CYCLE
x SERIES DURATION IN SECS.
COOLING TIME
MINUTES
10,000
20,000
35,000
60,000
130,000
1.0
2.0
3.0
4.0
5.0
FOR TRAINING PURPOSES ONLY!