5-12 Solar 8000M/i patient monitor 2026265-075C
Troubleshooting: Troubleshooting procedure
Troubleshooting procedure
Many network problems at the central station are traced back to the
patient monitor or Tram-rac housing. Therefore, many of the
troubleshooting suggestions in the “Problems and Solutions” section
include replacing pcb assemblies in the patient monitor or Tram-rac
housing. The “LED Troubleshooting Table” explains the normal
condition of the various LEDs on the pcb assemblies.
Do not perform the procedures listed unless you are a trained field or
biomedical engineer. Refer to “Disassembly” on page 7-7 for circuit board
functions. To aid in disassembly of the unit, refer to “Disassembly” on
page 7-7.
WARNING
Patient data loss. This procedure is intended for use by
service personnel with advanced troubleshooting skills.
The consequences of the following steps may cause loss of
patient data and disruption of the entire Unity Network.