2026265-075C Solar 8000M/i patient monitor A-23
Appendix A – Technical specifications: Solar 8000M/i display
Non-medical grade displays
The Solar 8000M/i patient monitor with a non-medical grade display 
meets UL and IEC specifications if an isolation transformer is used, 
regardless of whether the computer-grade display meets the leakage 
current specification on its own.
The party assembling or modifying the medical electrical system is 
responsible to insure compliance with IEC 60601-1-1. Therefore, if GE 
installs a Solar 8000M/i system with a computer-grade display, GE is 
responsible for meeting the specification.
As a result GE will only install non-medical grade displays with 
appropriate isolation transformers. (See the following information.)
Isolation transformers
Powervar has designed an isolation transformer specifically for this 
application. Powervar headquarters (listed below) will process orders 
and drop-ship to any destination requested. When calling Powervar, 
identify yourself as a GE representative/customer to receive the GE 
partnership discount.
Powervar
28457 North Ballard Drive, Suite C
Lake Forest, Illinois 60045
Phone: 847-816-8585
Fax: 847-816-8988
Contact your local sales/service representative for part numbers and unit 
information.