2026265-075C Solar 8000M/i patient monitor 4-15
Maintenance and checkout: Electrical safety tests
NOTE
If either reading is greater than the appropriate specification
below, the device under test fails. Contact GE Technical Support.
10 μA (0.01 volts on the DMM) at 120 VAC using the test body.
20 μA (0.02 volts on the DMM) at 240 VAC using the test body.
50 μA (0.05 volts on the DMM) at 120-240 VAC using the ECG
cable.
NOTE
The 10 and 20 μA limits are based on internal design standards.
The 50 μA limit is common to all standards. AAMI ES-1 standard
requires using the patient cable.
6. Set the power switch on the device to OFF.
7. Repeat the steps in this procedure using the appropriate SpO2 Test
Body. Connect the SpO2 Test Body to the blue SpO2 connector of the
device under test.
Test completion
1. Disconnect the leakage tester from the power outlet.
2. Disconnect all test equipment from the device.
3. Disconnect the device power cord from the leakage tester.