4-18 Solar 8000M/i patient monitor 2026265-075C
Maintenance and checkout: Checkout procedure
Procedures
Solar 8000M/i Unity View display
NOTE
This check only applies to Solar 8000M systems with product code 3S
and 3T.
Refer to the specific manufacturer’s documentation.
Touchscreen
Verify that touching a parameter box displays the screen information for
that parameter.
Tram-rac housing
1. Verify that the power LED is ON at the Tram-rac housing.
2. Disconnect and reconnect the Tram-rac housing communication
cable. Verify the recovery of the waveforms.
3. If the Tram-rac housing has additional slots for input modules, insert
a BP module. Connect simulator and verify communication to the
monitor. Repeat for each slot.
4. If the Tram-rac housing has an optional power supply, check the
following on the connector that applies to your equipment.
Verify +16.5V is not present at pin 5 of the TRAM-NET
connector with respect to pin 9.
Verify +16.5V is not present at pin 5 with respect to chassis
ground of the Tram-rac housing.