6-8 Solar 8000M/i patient monitor 2026265-075C
Configuration: Configuring a patient monitor
Set country selection
Select DEFAULT, FRANCE, or GERMANY to choose a particular set of
GE factory defaults.
Activate the Boot Code:
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
4. Select 4 SET CONFIGURATION menu option.
5. In the Configuration Menu, select 4 Country Selection and
choose the desired set of factory defaults.
Set language
Follow the steps for Set Country Selection above, except in the
Configuration Menu, select 5 Set Language and select the desired
language.
WARNING
Changing the language will discharge the monitor and
erase any saved monitor defaults. The monitor defaults
will be set to the factory defaults.
Calibrate touchscreen
Access the CALIBRATE TOUCHSCREEN option, starting from the
MAIN menu.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the TRIM KNOB control to select the day and
month from monitor screen with leading zeros (e.g., July 4 = 0407).
3. Select CALIBRATE.
4. Select CALIBRATE TOUCHSCREEN.
5. Touch and hold each of the 4 Xs as they appear per on-screen
instructions.
Completion
The monitor is now ready for normal operation. At this time, perform the
“Checkout Procedure” found in Chapter 3, Maintenance.