3-10 Solar 8000M/i patient monitor 2026265-075C
Installation: Power up
Indicators
A green LED indicates that the unit is connected to an AC power source
and the power switch is turned on. There is a green/yellow LED above
each M-Port indicating the M-Port status.
Solid green indicates the device is communicating properly.
Slow flashing yellow indicates the device has been identified, but
there is no communication.
Quick flashing yellow indicates that too many identical devices are
connected or the device cannot be identified.
Refer to Chapter 5, “Troubleshooting” if an LED is not green.
Power up
NOTE
Check power voltage at your location and set power to either 120 V or
240V.
After making all connections, plug the power cord into an AC wall outlet,
turn the power switch to 1 (on), and turn on the display. The power LED
illuminates and after about 10 seconds a display appears.
If the Solar 8000M/i patient monitor does not work properly, refer to
Chapter 5, “Troubleshooting” .