EasyManua.ls Logo

GE Solar 8000M/i - Admit Menu

GE Solar 8000M/i
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-6 Solar 8000M/i patient monitor 2026265-075C
Configuration: Configuring a patient monitor
Admit Menu
The ADMIT MENU selection determines the function of the monitor.
This menu item is part of the SERVICE MODE menu.
Before programming the ADMIT MENU, you must know if the monitor
will be used for standard adult, neonatal, or operating room monitoring,
and if the monitor will be moved from room to room. All combinations are
explained below.
STANDARD configures the monitor to stay in one room for
stationary monitoring only. Monitors not connected to the Unity
network (Ethernet connection) must use STANDARD configuration
only.
ROVER configures the monitor to move from room to room for
stationary monitoring only.
COMBO configures the monitor to stay in one room for both
stationary and ambulatory (telemetry) monitoring. This monitor
displays all Tram module data combined with ECG data for
ambulatory patients.
ROVER COMBO configures the monitor to move from room to room
for both stationary and ambulatory (telemetry) monitoring.
1. Access ADMIT MENU option, starting from the MAIN menu. Select
MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the TRIM KNOB control to select the day and
month from monitor screen with leading zeros (e.g., July 4 = 0407).
3. Select MENU SETUP > ADMIT MENU.
4. Use the TRIM KNOB control to select the function of the monitor.
5. Press TRIM KNOB control to exit.
↓ ↑
027A

Table of Contents

Related product manuals