6-14 Solar 8000M/i patient monitor 2026265-075C
Configuration: Advanced user procedures
Change Ethernet address
The Ethernet address is an identification number assigned to each device
on the Unity Network. It must be done in Boot Code using a unique
password only if it has been corrupted.
WARNING
Lost Data. Duplication of an Ethernet address on a
network will cause lost data. If you change the factory
assigned Ethernet address, you must first record all other
Ethernet addresses used on your network to avoid
duplication.
Fax a password request to GE Customer Relationship Center at 877-633-
8181 to obtain a password. You must provide your product serial number
and Ethernet address. (The Ethernet address displays in the Boot Code
banner information.)
Activate the Boot Code program as follows:
1. Hold down NBP Go/Stop and Zero All on the keypad or remote
control.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
4. Select Option Menu > Change Ethernet Address.