2026265-075C Solar 8000M/i patient monitor 6-13
Configuration: Advanced user procedures
Troubleshooting monitor defaults transfer
Below is a list of error messages that may display when performing a
monitor default transfer.
Storing monitor defaults
Copying stored monitor defaults
Message Description
ERROR - UNIT DEFAULTS HAVE NOT
BEEN SAVED
Monitor defaults were not saved due to a
memory error in the monitor.
Message Description
ERROR COPYING UNIT DEFAULTS -
NETWORK ERROR
Either an error has occurred while the
defaults were being transferred, or defaults
have not been saved on the server monitor.
ERROR COPYING UNIT DEFAULTS -
UNIT TYPE MISMATCH
The server and client monitors are set to
different unit types (i.e., ADULT, NEO, or
OR)
ERROR COPYING UNIT DEFAULTS -
SOFTWARE VERSION MISMATCH
The server and client monitors have either
different revisions of software, or different
software versions (i.e., Solar 7/8000 vs.
Solar 8000i)
ERROR COPYING UNIT DEFAULTS -
DATA CORRUPTION ERROR
The monitor defaults transferred have
become corrupted, possibly due to a
memory error.
ERROR COPYING UNIT DEFAULTS -
SOFTWARE FEATURE MISMATCH
The server and client monitors are set to
different software levels (i.e., Basic,
Cardiac).
ERROR COPYING UNIT DEFAULTS -
COUNTRY CODE MISMATCH
The server and client monitors are set to
different country codes (i.e., DEFAULT,