4-16 Solar 8000M/i patient monitor 2026265-075C
Maintenance and checkout: Checkout procedure
Checkout procedure
General
These procedures test the functions of the patient monitor, Tram-rac
housing or Patient Data Module and associated communication
networks. For other input modules checkout procedures, refer to their
appropriate service manual.
GE recommends that all checkout procedures be performed
upon receipt of the device,
once a year thereafter,
after any upgrade,
any time the main enclosure is disassembled, or a part is repaired or
replaced.
Record the date and results on the Repair Log included at the end of this
chapter.
Required tools and equipment
NOTE
Be sure test equipment is maintained according to manufacturer’s
maintenance schedule to ensure accuracy.
The following table lists the test equipment, adapters, and cables
necessary to complete the checkout procedures. Equivalent equipment
may be substituted
Acquisition device:
Tram 100-851 module (Any GE Tram 100-851 module)
Patient Data Module (Any GE Patient Data Module)
Multifunctional patient simulator (MARQ III or equivalent)
Patient cables (Multi-link ECG 12-lead patient cable and ECG leadwire set)
Tram-rac housing test:
Tram-rac (Any GE Tram-rac)
BP module (Any GE BP module)
Digital multimeter (AC volts, ohms)
M-Port test:
Port Checkout DIDCA (420915-031)
Cable, category 5 (418335-002)
AutoPort to M-Port Adapter (2001973-001)
Blood pressure test:
Blood pressure simulator cable (700095-001)