5-2 Solar 8000M/i patient monitor 2026265-075C
Troubleshooting: Terms used
Terms used
Listed and defined below are terms used in this chapter.
Abort (main code)
This is a Main Code menu selection that may appear on the monitor
during software downloads from the monitor SERVICE MODE menu.
ABORT causes the download to stop by pushing the TRIM KNOB
control.
Boot loader or boot code
The Boot Loader or Boot Code is used to download software. Entering the
Boot Code does not erase any memory, but downloading new software
will erase protected memory. To activate the Boot Code, perform the
following.
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
Cold start
A cold start is used only in extreme circumstances. It erases the
protected memory, language, unit name and bed number; automatically
discharges the patient; and restores factory defaults. Ethernet address,
internet address, software feature level, and CRG trend option remain
unchanged.
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control and keep holding NBP
Go/Stop and Zero All.
3. When the monitor resets and the display goes blank, press and hold
Display On/Off (in addition to NBP Go/Stop and Zero All) until
the message “performing cold start” appears on the display.
NOTE
If the Boot Code information appears before the message
performing cold start” appears, begin the cold start again.
Continue (main code)
CONTINUE is a Main Code menu selection that appears after a
successful software download. It allows the user to continue downloading
other files without resetting the monitor.