2026265-075C Solar 8000M/i patient monitor 5-3
Troubleshooting: Terms used
Monitor memory
Monitor memory contains the Ethernet address, internet address, bed
name, care unit name, and standard unit defaults for each monitor.
Protected memory
Protected memory contains a patient’s history and any individualized
changes to the unit defaults.
Power cycle or reboot
To power cycle or reboot the monitor, turn the power OFF at the rear of
the unit, wait 10 seconds, then turn the power ON.
Service mode (main code)
The SERVICE MODE menu is found in the monitor MAIN menu and is
used for various functions like calibration, video tests, and downloading
monitor interface software. (See details later in this chapter.)
Service menu (boot code)
The Service Menu found in the Boot Code is used when downloading the
Boot Code and main processor code. (See details later in this chapter.)
Warm start (boot code)
A warm start activates the software previously downloaded. The
following steps activates a warm start.
1. Hold down NBP Go/Stop and Zero All.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
4. Select Start Patient Monitoring.
Country selection
This is a Boot Code setting. The choices are France, Germany, or
Default which chooses a particular set of GE factory defaults. These
defaults are used when changing the unit type (Adult-ICU, OR, or
Neonatal).
Set language
This is a Boot Code setting. It sets the language for displayed text in the
Main Code. (Do not confuse with Country Selection.)