Appendix A – Technical specifications: ICG module
ICG module
Performance specifications
Measurement Range
Cardiac Output (CO):
Cardiac Index (CI):
Stroke Volume (SV):
Stroke Index (SI):
Acceleration Index (ACI):
Thoracic Fluid Content (TFC):
Pre-Ejection Period (PEP):
Left Ventricular Ejection Time (LVET):
Systolic Time Ratio (STR):
Velocity Index (VI):
Systemic Vascular Resistance (SVR):
Indexed Systemic Vascular Resistance
(SVRI):
Left Stroke Work Index (LSWI):
Indexed Left Cardiac Work (LCWI):
Heart Rate (HR):
0 to 30 L/min
0 to 15 L/min/m
2
0 to 250 mL
0 to 125 mL/m
2
0 to 400 1/100s
2
10 to 150 1/kΩ
0 to 1000 milliseconds
0 to 1500 milliseconds
0 to 1
0 to 200 1/1000s
0 to 5000 dyne
.
sec
.
cm
-5
0 to 10,000 dyne
.
sec
.
cm
-5
/m
2
0 to 200 g
.
m/m
2
0 to 25 kg
.
m/m
2
40 to 250 beats/min
Power requirements
Maximum Power Consumption:
(non-isolated)
+16.5 volts, 610 mA
+5.0 volts, 135 mA
Environmental specifications
Operating Conditions Refer to the monitor specifications.
Storage Conditions
Temperature:
Relative Humidity:
–40°C to +70°C (–40°F to +160°F)
0 to 95% (non-condensing)
Altitude: -305 to 3353m (-1000 to 11,000ft.)
Atmospheric Pressure: 645 to 795mmHg (860 to 1060hPa)
Physical specifications
Height: 4.0 cm (1.6 in)
Width: 11.4 cm (4.5 in)
Depth: 28.6 cm (11.25 in)
Weight: 1.0 kg (2.2 lb) maximum
Cooling Method: Forced Air
Heat Dissipation: 9W maximum
Compatible host monitors, Tram-racs
Host Monitor/Tram-rac Software Version
Solar 7000: 7A or later
Solar 8000: 7A or later
Tram-rac 4A: 6A or later
Warranty One year (accessories may differ)