5-6 Solar 8000M/i patient monitor 2026265-075C
Troubleshooting: Service menus
Main code SERVICE MODE menu
Access the SERVICE MODE menu starting from the MAIN menu.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the TRIM KNOB control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407).
DOWNLOAD CODE
This menu provides options for downloading SpO
2
, RAC COMM/DAS,
SAM, and CO
2
modules.
CALIBRATE
These menu selections are used to calibrate the touchscreen and various
inputs from the Tram module, Patient Data Module, or other discrete
modules. Refer to the module service manual for details. See Chapter 6,
“Configuration” , for calibration details.
HARDWARE TEST
This menu provides access to the hardware tests for the monitor.
Hardware Status
The ‘one of a kind’ patient monitor LAN address is listed here. The
current revision of the main processor and FPGA are also listed.
Video Test
WARNING
Loss of patient data. Do not enter this menu selection
unless loss of patient data is not an issue.
Provides a list of various test patterns and colors that can be used to
calibrate the display. When finished, select RESTART System.
PATIENT-MONITOR TYPE
Select the type of monitor desired, i.e., adult, neonatal or operating room.
See Chapter 6, “Configuration” , for detailed procedures.