2026265-075C Solar 8000M/i patient monitor 5-5
Troubleshooting: Service menus
Clear Monitor Memory
If you respond with YES in this option, you will erase all patient 
histories and programmed defaults in the protected memory and return 
to factory installed defaults.
Set Configuration
This menu contains options for defib sync for ECG modules (not Tram 
modules), line frequency, country selection (not for text translations), 
and setting language for displayed text.
WARNING
Changing the language will discharge the monitor and 
erase any saved monitor defaults. The monitor defaults 
will be set to the factory defaults.
Serial Download Main
This option is used when downloading software from a laptop PC.
Serial Download Boot
This option is used when downloading software from a laptop PC.
View Main Code Error Logs
Allows access to the error logs generated by the Main Code.
Tools Menu
This password protected menu contains service diagnostic tools. Use 
password āmei^ā.
Video Test Screens
Various color screens for testing the display.
Options Menu
In this menu, you can enable various software options (Hires Trends, 
Cardiopulmonary Features, etc.) and change the monitor software level. 
A unique password is required for each option and to change the software 
level or Ethernet address. Fax a password request to GE Customer 
Relationship Center at (414) 362-3250 to obtain a password. You must 
provide your product serial number and Ethernet address. (The Ethernet 
address displays in the Boot Code banner information.)