7-8 Solar 8000M/i patient monitor 2026265-075C
Field replaceable units: Disassembly
Hardware
Before disassembly, note the positions of any wires or looms (cables),
marking them if necessary to ensure that they are replaced correctly.
Gray ribbon cables have retainer clips holding them in the connector.
Save and set aside all hardware for reassembly.
Disassembly procedures
WARNING
Patient monitoring interruption. Make sure a patient is
not being monitored.
First, turn the unit off at the rear power switch and disconnect the AC
power cord and all communication cables.
WARNING
Due to possible high voltage present, use an insulated
screwdriver at all times.
Refer to the exploded view of the unit in this chapter when
disassembling. Reassemble the unit in reverse order of disassembly
unless otherwise stated.
WARNING
Shock hazard. High voltages exist in this unit. Use
insulated tools. Remove jewelry. Use only one hand when
possible.
Cover
1. Remove four painted screws on the sides of the unit.
NOTE
Use the #1 Phillips head screwdriver.
2. Slide the cover directly back toward the rear to clear the right angle
piece of the M-port ethernet cable, then lift cover off.