6-2 Solar 8000M/i patient monitor 2026265-075C
Configuration: Configuring a patient monitor
Configuring a patient monitor
The following procedure explains how to configure a patient monitor on
the Unity Network. The monitor communicates with central stations,
and other related equipment over the Unity Network. This network is
essentially an Ethernet implementation.
General
Use this procedure if you are:
experiencing communication problems on the Unity Network, or
adding a new monitor to the Unity Network.
Gather information
To configure a new monitor, you must first:
know that the new monitor’s software revision is compatible with the
other monitors connected to the Unity Network.
write down the exact care unit name from the upper left hand corner
of the central station.
write down the bed name for the new monitor.
know if the monitor will be used for either stationary or ambulatory
(telemetry) monitoring or both.
know if the monitor will be moved from one Ethernet connection to
another.
Select procedures
Choose and program the procedures listed below in the order presented.
Each procedure is described on the next pages.
Set Unit Name (Main Code)
Set Bed Number (Main Code)
Patient-Monitor Type (Main Code)
Set Graph Locations
Admit Menu (Main Code)
Set Line Frequency (optional, Boot Code)
Set Defib Sync Voltage and Pulse Width (optional, Boot Code)
Set Country Selection (Boot Code)
Set Language (Boot Code)
Calibrate Touchscreen
After completing all necessary procedures, perform the “Checkout
procedure” on page 4-16.