6-10 Solar 8000M/i patient monitor 2026265-075C
Configuration: Advanced user procedures
Change software level
If you want to change the software to a higher level or add software
options, it must be done in Boot Code using a unique password.
Fax a password request to GE Customer Relationship Center at (414)
362-3250 to obtain a password. You must provide your product serial
number and Ethernet address. (The Ethernet address displays in the
Boot Code banner information.)
Activate the Boot Code program as follows:
1. Hold down NBP Go/Stop and Zero All on the keypad or remote
control.
2. Press and release the TRIM KNOB control.
3. Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
4. Select Option Menu > Change Software Level.
Enable options
For information about enabling software options see, “Service menu (boot
code)” on page 5-3.
Transfer monitor defaults
The monitor defaults set on one monitor may be transferred to another
monitor, provided that the second monitor shares the following:
the same software revision (version),
the same patient-monitor type (i.e., ADULT, NEO, or OR),
the same software package (i.e., Basic or Cardiac), and
the same country code (i.e. DEFAULT, FRANCE, or GERMANY).
The following defaults transfer when using this feature:
all monitor defaults
custom default names
12 SL location
12 SL site