2026265-075C Solar 8000M/i patient monitor 6-11
Configuration: Advanced user procedures
Store monitor defaults for transfer
To transfer monitor defaults, they must be set, then stored on the 
monitor that is used as the server.
1. At the server monitor, set up the monitor defaults you want to store. 
(Refer to Monitor Setup in the Solar 8000i Operator’s Manual.)
2. When defaults are set, start from the MAIN menu and select MORE 
MENUS > MONITOR SETUP > SERVICE MODE.
3. Enter password using the TRIM KNOB control to select the day and 
month from monitor screen with leading zeros (e.g., July 4 = 0407).
4. Select MONITOR SETTINGS.
5. Select STORE DEFAULTS FOR NETWORK TRANSFER.
6. Selects YES to the confirmation popup menu.