2026265-075C Solar 8000M/i patient monitor 5-7
Troubleshooting: Service menus
WARNING
Changing the patient-monitor type will default the admit
function to STANDARD configuration. Different alarms
and parameters are activated for each selection.
NOTE
The keypad/remote control is DIDCA programmed for specific
monitor types. The error message WARNING: REMOTE
MISMATCHED WITH MONITORING MODE displays if the
monitor and keypad/remote control do not match.
MENU SETUP
This menu selection provides the following submenus: (See Chapter 6,
“Configuration” , for detailed procedures.)
ADMIT MENU: STANDARD
This menu selection allows you to determine the function of the patient
monitor. The four variables include stationary or ambulatory (telemetry)
patient monitoring with a monitor that always stays in one room
(STANDARD) or a monitor that moves from room to room (ROVER).
SOFTWARE LEVEL
This menu selection displays the software feature level this monitor is
using. It allows setting the level to a lower setting than the software
feature level setting in Boot Code.
MONITOR DEFAULTS PASSWD
This menu selection is used to turn a password requirement ON or OFF
for entry into the MONITOR DEFAULTS menu section described
above. If selected, the password will be the same as the SERVICE
MODE MENU password.