2026265-075C Solar 8000M/i patient monitor 4-19
Maintenance and checkout: Checkout procedure
5. The following step does not apply for a Tram-rac 2 housing. Check
the analog output connector (yellow) using an oscilloscope. Observe a
signal at the appropriate pins found in the next table. The output
signal is dependent upon which Tram and input module functions
are activated at the monitor.
12
6
1
11
10 7
9 8
2
5
4
3
REAR VIEW
TRAM-RAC 4A
ANALOG OUTPUT
CONNECTOR
FRONT VIEW
TRAM-RAC 3
ANALOG OUTPUT
CONNECTOR
8
7
6
5
4
3
2
1
14
13
12
11
10
9
15
019A
Analog output signals
Pins for
D-type
connector
Pins for round
connector
Signal source
Tram-rac 4A
bezel number for
BP output
Pin 1 Pin 8 Signal GND for Tram Waveforms
Pin 2 Pin 2
Trace I (ECG II
1
)Tram
1
Pin3 Pin 6 Tram BP3 or SPO
2
Value
Tram
1
Pin 4 Reserved for Future Use
Pin 5 Pin 4 Tram ART 1 or BP1 Tram
Pin 6 Pin 9 Slot 3 Series 7000 Waveform A (Right Side or Module) Parameter 6
Pin 7 Pin 11 Slot 4 Series 7000 Waveform A (Right Side or Module) Parameter 8
Pin 8 Pin 8 Signal GND for Series 7000 Waveforms
Pin 9 Pin 1 Tram ECG II
Tram
1
Pin 10 Pin 3 Tram ECG V
Tram
1
Pin 11 Pin 7 Tram BP4 or RESP
Tram
1
Pin 12 Reserved for Future Use
Pin 13 Pin 5 Tram BP2 or SPO
2
Waveform Tram
Pin 14 Pin 10 Slot 3 Series 7000 Waveform B (Left Side or Module) Parameter 5
Pin 15 Slot 4 Series 7000 Waveform B (Left Side or Module) Parameter 7
1. The top displayed trace on the monitor is present unless AVR, AVL, or AVF leads are used,
then lead II is output.