5-24 Solar 8000M/i patient monitor 2026265-075C
Troubleshooting: Reviewing error/event logs
For some categories of TRAM-NET network errors, two additional lines
are added to the error log entry.
Severity is a measure of how the event/error affected the system. There
are three levels of severity.
Definition of parameters
Name Description
Network Error Identifies that a network error occurred.
Channel Number Identifies the network channel exhibiting the error.
Definition of severity levels
Levels Description
Continue The event or error was logged, the task may have or may not
have been finished, but the system was able to go on. Most log
entries will have a severity of CONTINUE.
Fatal The event or error was such that the task is not able to go on.
Recovery was not possible. This always is followed by a warm
start.
Forced Restart The system was restarted by a known condition (internet
address change, video test, etc).