5-26 Solar 8000M/i patient monitor 2026265-075C
Troubleshooting: Reviewing error/event logs
Useful event data
The event log tracks alarm events and other changes to the system that
can be useful in determining the state of the monitor if an error occurs.
For each event, the event log records the event number, the date, the
time, the event ID, the “bucket,” and an event description. Additionally,
an event description code is logged.
NOTE
“Bucket” data is information regarding the state of the event. For
example, if the monitor’s automatic alarm graphing is turned on or
off, the event bucket will contain the number 0 or 1, which
correspond to off and on.
The table below shows some common events, their event codes, the
corresponding bucket data (if applicable) and other event data (if
applicable). This data can be useful in interpreting the event log.
Event description Event ID Bucket data Other event data
Alarm Graph On/Off Change 800 0 = Alarm graphing turned OFF
1 = Alarm graphing turned ON
Graph Location Change 801 1 = Manual graph location changed
2 = Alarm graph location changed
3 = Print window location changed
4 = 12 lead print location changed
Text describing the graph name is
shown.
Monitor Defaults Change 802 Text describing the default changed
is shown (e.g., COLOR FORMAT).
Alarm Limits Change PPtt 803 Limit is indicated in hex.
Viewing Another Bedside On 804 1 = Split View
2 = Full View
Text describing the viewed bedside
name is shown.
Viewing Another Bedside Off 805 1 = Split View
2 = Full View
Text describing the viewed bedside
name is shown.
Network Connection Change 806 0 = Stops communicating
1 = Starts communicating
TRAM/ECG Module Connection Change 807 0 = Stops communicating
1 = Starts communicating