2026265-075C Solar 8000M/i patient monitor 7-11
Field replaceable units: Disassembly
Lithium battery
The following error message will appear when the lithium battery fails.
ERROR: THE INTERNAL BATTERY THAT MAINTAINS THE 
MONITOR’S DEFAULTS HAS FAILED!”
* * SERVICE MAY BE REQUIRED * *
When this error message appears, complete the steps listed above for the 
processor board. Then desolder the battery and solder in a new lithium 
battery.
Speaker
1. Remove the cover.
2. Using a metric hex or allen wrench with ball end, unscrew two 
screws.
3. Remove speaker from chassis.