A-2 Solar 8000M/i patient monitor 2026265-075C
Appendix A – Technical specifications: Solar 8000M/i patient monitor
Solar 8000M/i patient monitor
Due to continual product innovation, specifications and standards are subject to change without notice. The
following specifications are accurate as of the date of this publication, and pertain to Solar 8000M/i patient
monitor.
Specifications are based on a complete system including a 15 inch monitor (display*).
Performance specifications
Solar 8000M/i processing unit controls
Optional keypad with Trim Knob control and 18 active hard keys. See Keypad for applications and specifications.
Optional remote control with Trim Knob control and 18 active hard keys. See Remote Control for applications and specifications.
Optional touchscreen interface: Requires touchscreen display and software.
Communications protocols:
TRAM-NET Proprietary CSMA/CD protocol at 921 Kbits/sec
Unity Ethernet IEEE 802.3 10 BASE T LAN
Processing
Main processor: Motorola
®
MPC860P (PowerPC), 66 MHz, 32 bit
Graphics processor: Silicon MotionGraphics Accelerator with Silicon Image Transmitter providing
images on VGA and DFP displays
TRAM-NET communication processor: Within Motorola MPC860P (PowerPC), 66 MHz, 32 bit
LAN communication processor: Within Motorola MPC860P (PowerPC), 66 MHz, 32 bit
Alarms
Classification:
Patient Status Alarms — 4 levels: Crisis, Warning, Advisory, and Message
System Status Alarms — 2 levels: Warning and Advisory
Notification: Audible and visual, dependent on alarm level
Display of alarm information: All limits are viewable and printable
Continuous display of limits: All parameters, one set of limits
Interfaces
Contact your local sales representative.
Environmental specifications
Power requirements: 110 ±20 VAC, 50/60 Hz, single phase
230 ±40 VAC, 50/60 Hz, single phase
Power consumption: 100 Watts
Low-voltage shutdown: <90 VAC/<190 VAC
Cooling: Natural convection
Heat dissipation: 341 Btu/hr