A-8 Solar 8000M/i patient monitor 2026265-075C
Appendix A – Technical specifications: Tram modules and Solar parameter functionality
Analog output: Selectable at 1V/mV
Defibrillator sync/analog out delay: <35 ms
Display:
Standard mode: 0.05 to 40 Hz
OR Mode: 0.05 to 25 Hz
Neonatal Mode: 0.5 to 40 Hz
Laser Printer:
Standard Mode: 0.05 to 100 Hz
OR Mode: 0.05 to 25 Hz
Neonatal: 0.5 to 40 Hz
Heart rate:
Heart rate averaging: 8 Beats
Heart rate accuracy: ± 1% or ± 1 beat/minute, whichever is greater
Display update interval: 2 seconds
Response time: < 6 seconds (per AAMI EC13)
Limit alarm delay: < 10 seconds after limit alarm condition exceeded
Heart rate alarm range: -1 to 300 beats/minute, high limit > low limit
Arrhythmia analysis: 1 to 100 PVCs/minute
Method: QRS morphology classification and timing based on single- or multiple-lead analysis
Arrhythmia calls: Full, lethal only, or no arrhythmia
PVC alarm limit range: 1 to 100 PVCs/minute
ST Segment analysis:
Measurement description: ST segment deviation is measured and displayed for all acquired leads
ST display: Lead label, ST deviation, current complex superimposed over a reference complex, J-
point indicator and 15-minute mini-trends are shown for all acquired leads
Measurement point: Measured at user-selectable measurement points (0, 30, 40, 50, 60, and 80 ms) following
the J point
Measurement range: -12.0 mm to + 12.0 mm
Display resolution: 0.1 mm
ST measurement averaging: 16 beats
Display update interval: 2 seconds
ST alarm limits: ± 12 mm, high limit > low limit, for any event within a lead group (inferior, lateral or
anterior) that exceeds the alarm limit for that group
Pacemaker detection/rejection:
Input voltage range ±2 mV to ±700 mV
Input pulse width 0.1 ms to 2 ms
Rise time 10 μs to 100 μs
Over/under shoot 2 mV (maximum)