A-12 Solar 8000M/i patient monitor 2026265-075C
Appendix A – Technical specifications: Tram modules and Solar parameter functionality
Pressure accuracy:
Static:
Clinical:
±2% or ±3 mmHg, whichever is greater
±5 mmHg average error
8 mmHg standard deviation
Measurement modes: Manual, auto, and stat
Heart rate detection: 30 to 200 beats per minute
Total cycle time: 20 to 40 seconds typical (dependent on heart rate and motion artifact)
Maximum inflation pressure: 300 mmHg (adult), 250 mmHg (pediatric), 150 mmHg (neonatal)
Over-pressure valve: Activates when cuff pressure exceeds:
300 mmHg (+30/-0 mmHg) adult,
150 mmHg (+15/-0 mmHg) neonatal
Automatic cycle times: 0 to 24 hours
Auto zero: Zero pressure reference prior to each cuff inflation
Tubing length: 12 feet adult, 8 feet neonatal
Automatic cuff deflation: Cycle time exceeding 2 minutes (85 seconds neonatal), power off, or cuff pressure
exceeds 290 mmHg (± 6 mmHg) for adult, 250 mmHg (± 6 mmHg) for pediatric, or 145
mmHg (± 5 mmHg) for neonatal
Cuff sizes:
Disposable: Large adult, adult, small adult, pediatric, child, and neonatal
Reusable: Adult thigh, large adult, adult, small adult, small adult/child, child, and infant
Alarms: User-selectable upper and lower limits for systolic, diastolic, and mean pressures
Pulse oximetry
Tram 200SL, 400SL, 600SL, 800A, 800SL and 300 (support Ohmeda probes)
Tram 250SL, 450SL, 650SL, 850A and 850SL (support Nellcor and GE probes)
Parameters monitored: Arterial oxygen saturation (SpO2) and peripheral pulse rate (PPR)
Probe types: GE, Nellcor, Ohmeda.
Range:
SpO2: 0 to 100%
PPR: 20 to 300 beats per minute (±3 beats per minute)