3-8 Solar 8000M/i patient monitor 2026265-075C
Installation: Front panel connectors and indicators
The following devices connect to M-Port hosts using AutoPort to M-Port
adapter PN 2001973-001.
PRN-50 digital writer with AutoPort
Respiratory mechanics module with AutoPort
NOTE
AutoPort to M-Port adapter PN 2001973-001 is required for
connecting AutoPort devices to M-Ports. Plug the adapter end
labeled AutoPort into the AutoPort device.
Remote control or keypad
The remote control or keypad is DIDCA programmed for specific care
areas (adult, neonatal, or operating room). A keypad holster mount is
available for mounting under the display.
NOTE
The Solar 8000M/i requires an interface (remote control or keypad)
with all display types.
NOTE
The error message WARNING: REMOTE MISMATCHED WITH
MONITORING MODE displays if a mismatched keypad/remote
control is connected to the Solar 8000M/i patient monitor.
PRN-50 digital writer and respiratory mechanics module
The figure below shows the PRN-50-M digital writer connected to one of
the M-Ports. The RM-M respiratory mechanics module has similar
connections.
AutoPort or M-Port connector.
Note: If this is a PRN 50 or RM module with
AutoPort, then adapter PN 2001973-001 is
required.
016A