2026265-075C Solar 8000M/i patient monitor 4-7
Maintenance and checkout: Cleaning
Improper cleaning products and processes impact/results
Product discoloration.
Metal part corrosion.
Brittle wires.
Brittle and breaking connectors.
Reduced cables and leadwires life.
Unit malfunction.
Void warranty.
Cleaning products to avoid
Cleaning products known to cause the types of problems listed above
include, but are not limited to:
Sani-Cloth
®
Wipes
Ascepti
®
Wipes
HB Quat
Clorox
®
Wipes (they do not contain bleach).
Over-the-counter detergents (e.g. Fantastic
®
, Tilex
®
, etc.).
Products that contain active ingredients and solutions similar to these
products should also be avoided.
NOTE
For additional information, refer to the How to Reach Us page of the
manual for contact information. Also see the GE Handheld Medical
Devices Cleaning, Disinfecting, and Storage addendum.