2026265-075C Solar 8000M/i patient monitor 4-21
Maintenance and checkout: Checkout procedure
Keypad/remote control
1. Plug the keypad or remote control into an M-port.
2. Check all functions of the TRIM KNOB control and 18 hard keys.
Verify a response at the monitor display.
3. Check that the backlight is on and lights the keys evenly.
4. Activate Boot Code as follows:
Hold down NBP Go/Stop and Zero All.
Press and release the TRIM KNOB control.
Keep holding NBP Go/Stop and Zero All until the Boot Code
information appears on the display.
5. Select Tools Menu.
6. Use password “mei^” to open Service and Diagnostic Tools
Menu.
7. Select M-Port Tools > Show M-Port #X Serial Incoming Data.
(Where X = the M-Port number.)
8. If the message, “No incoming data read” appears, nothing is
connected to the M-Port. Connect the keypad or remote control or
choose the correct M-Port.
9. As the columns of numbers scroll, press the keys and TRIM KNOB
control. Verify that the numbers change as keys are pressed.
10. Exit Boot Code.
Unity Network ID
Unity Network ID connectivity device checkout procedures, see the Unity
Network Interface Device (ID) service manual.
Printer
Press Graph Go/Stop and verify that the printer responds correctly.