4-34 Solar 8000M/i patient monitor 2026265-075C
Maintenance and checkout: Checkout procedure
NOTE
The Marker Out amplitude and pulse width are configured at 5V and
10ms respectively from the factory. To change these settings refer to
the Calibration chapter of your TRAM modules service manual.
ECG
Signal Pin: 7
Ground Pin: 3
Probe Type: x10
Time/Division: 0.2S
Volts/Division: 0.5V
Arterial BP
Signal Pin: 6
Ground Pin: 5
Probe Type: x10
Time/Division: 0.2S
Volts/Division: 0.2V
023A
024A