600 Series Steam Sterilizers
INS 61301607027 Rev A US 5–23
FIGURE 5–19. STERILIZER PIPING WITH WATER-SAVER INSTALLED
NOTE
The Water-Saver requires a separate 115 Vac, 60 Hz (26A rating) or 230 Vac,
60 Hz (13A rating) mains supply.
A08205-F
(30–60 PSIG)
WATER
SUPPLY
2ST
6SV
5SV
2SV
15SV
AI-04
5RT
DRAIN
5CV
9CV
DO-06
CE RE
INSTRUMENTATION DIAGRAM
W/EJECTOR VACUUM & WATER SAVER
JKT JKT JKT
JKT JKT JKT
CHAM
VALIDATION PORT
DOOR SEAL A
DOOR SIDE A
DOOR SEAL B
DOOR SIDE B
CHAMBER
NOTES:
REFER TO CORRESPONDING PROCESS
DIAGRAM (REFERENCED IN THE UNIT’S
SOFTWARE PACKAGE) FOR CYCLE PHASE
DIAGRAMS AND PROCESS DESCRIPTION.
CE = CONTROL END (SIDE A)
RE = REMOTE END (SIDE B)
1)
2)
WATER SAVER
EXISTING COMPONENTS
ON STANDARD P&ID’s
EXISTING CONNECTIONS
ON STANDARD P&ID’s
SIGNAL TO DEVICE
SOLENOID VALVE
(NORMALLY CLOSED
TWO-WAY VALVE)
CHECK VALVE
STRAINER
EJECTOR
LEGEND