Alar
m
Alarm text Description
ET33 A33 CONFIG ERROR Bad configuration. May occur when reconfiguring the cycle.
Possible action:
There is a set of configuration combinations that is not allowed in this
system. (parameter flags PF vs. DIP contacts DS). Make sure none of the
combinations in the A33 Config list are valid. See sectionDescription of
A33 alarm (configuration alarm), Page 119
ET34 A34 I/O FAULT Fault 12C bus communication or fault I/O card.
Possible action:
a. Check the connections for the 12C bus.
b. Check the connections on the analogue sensors.
c. Check the fuses for the analogue sensors.
ET35 A35 MACHINE STOP Emergency stop on soiled side, clean side or in service area has been
activated.
Possible action:
a. Check why someone have activated the emergency stop
button.
b. If no danger reset the machine stop button.
ET36 A36 DRYER TEMP H If the temp of the drying air >140°C.
Possible action:
a. Check that the fan is working.
b. Check that the steam valve closes.
c. Check the calibration of the temperature sensor.
ET37 A37 SAFETY LINE Safety line inside the washer activated or adjusted too loosely.
Possible action:
a. Check why someone have activated the safety line.
b. Check the installation of the safety line.
c. If no danger reset the safety line.
ET38 A38 CHAMB TEMP H If the chamber temperature is more than set point +5°C during wash 1,
final wash or final rinse.
Possible action:
a. Check temperature set point compared to temperature for
saved water.
b. Check the parameters for the temperature control.
c. Check steam pressure and steam valve.
ET39 A39 T SENS DRAIN If broken sensor temperature drain.
Possible action:
a. Check the connection of the temperature sensor (for damages
or short circuit).
ET40 A40 DRAIN TEMP H Temperature draining water >60°C for more than 15 sec.
Possible action:
a. Check cooling water supply.
b. Check cooling water valve function.
c. Check cooling water temp.
Page 115 of 166