EasyManua.ls Logo

Giga-tronics 58542 - Setup Parameters; Equipment Required; Figure 4-3: Power Sensor Linearity Test Setup

Default Icon
252 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration & Testing
Publication 21555, Rev. E, September 2002 4-11
4.4.2.1 Setup Parameters
The following setup parameters should be accomplished prior to performing the Power Linearity test:
4.4.2.2 Equipment Required
GPIB Controller · Digital Voltmeter · RF Signal Generator · Thermistor Power Meter · Directional Coupler · Step
Attenuator · Power Sensor
1. The power meter and sensor should be calibrated by following the instructions in Section4.3 of this
publication.
2. The Averaging is set to AUTO by entering:
SENS1:AVER:COUN:AUTO ON
(Enter SENS2 for channel 2)
Extreme care is required in the following procedure since the ac-
curacy requirements are critical to ensure the most accurate test
results.
Figure 4-3: Power Sensor Linearity Test Setup
CAUTION

Table of Contents

Related product manuals