Introduction
Publication 21555, Rev. E, September 2002 1-5
1.2.3 Zeroing Accuracy (Standard CW Sensors)
Notes:
1. Depending on sensor used.
2. Specified performance applies with maximum averaging and 24-hour warm-up at constant temperature.
System Linearity (@ 50 MHz for Standard CW Sensors)
±0.02 dB over any 20 dB range from -70 - +16 dBm
±0.02 dB + (0 dB, -0.05 dB/dB) from +16 - +20 dBm
±0.04 dB from -70 - +16 dBm
Linearity Temperature Coefficient
<0.1%/ °C temperature change following Power Sweep Calibration, 24-hour warm-up required
<0.3%/ °C temperature change following Power Sweep Calibration, 24-hour warm-up required (8035XA Series Sensors)
Zero Set
<±50 pW
2
Zero Drift
<±100 pW during 1 hour
2
Noise
<±50 pW measured over any 1 minute interval
2
Averaging
Auto-averaging or user-selectable averaging from 1 to 512 readings per measurement