EasyManua.ls Logo

Giga-tronics 58542 - Page 191

Default Icon
252 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Program Examples
Publication 21555, Rev. E, September 2002 A-9
-300, Device-specific error;No Sensor
Calibration FAILED, sensor on calibrator?
410 ASSIGN @Pwr_mtr to 70101
420 CLEAR @Pwr_mtr
440 WAIT 1
450 INPUT Disconnect the sensor from the cable 1 second after hitting ENTER. Hit ENTER.,Dmy
460 PRINT Calibrating Sensor 1
470 OUTPUT @Pwr_mtr;CAL1?
480 ENTER @Pwr_mtr;Pass_cal
490 IF Pass_cal=0 THEN PRINT Calibration Passed
500 IF Pass_cal=1 THEN
510 Ques$=Calibration FAILED, sensor on calibrator?
520 OUTPUT @Pwr_mtr;SYST:ERR?
530 ENTER @Pwr_mtr;Err_msg$ ! Reading an error message clears it.
540 PRINT Err_msg$
550 PRINT Ques$
560 ELSE
570 IF Pass_cal<>0 THEN PRINT Strange Response to CAL1?. Clear output Queue?
580 END IF
590 !
600 !
610 STOP
Program Output:
Calibrating Sensor 1
-300, Device-specific error;Sensor not connected to calibrator
Calibration FAILED, sensor on calibrator?
410 ASSIGN @Pwr_mtr to 70101
420 CLEAR @Pwr_mtr
440 WAIT 1
450 INPUT Do not connect the sensor to the calibrator. Hit ENTER.,Dmy
460 PRINT Calibrating Sensor 1
470 OUTPUT @Pwr_mtr;CAL1?
480 ENTER @Pwr_mtr;Pass_cal
490 IF Pass_cal=0 THEN PRINT Calibration Passed
500 IF Pass_cal=1 THEN
510 Ques$=Calibration FAILED, sensor on calibrator?
520 OUTPUT @Pwr_mtr;SYST:ERR?
530 ENTER @Pwr_mtr;Err_msg$ ! Reading an error message clears it.
540 PRINT Err_msg$
550 PRINT Ques$
560 ELSE
570 IF Pass_cal<>0 THEN PRINT Strange Response to CAL1?. Clear output Queue?
580 END IF
590 !
600 !
610 STOP
Program Output:
Calibrating Sensor 1
-300, Device-specific error;Sensor not connected to calibrator
Calibration FAILED, sensor on calibrator?
410 ASSIGN @Pwr_mtr to 70101
420 CLEAR @Pwr_mtr
440 WAIT 1
449 PRINT Connect the sensor to the calibrator.
450 INPUT Remove the sensor 10 seconds after hitting ENTER. Hit ENTER.,Dmy
460 PRINT Calibrating Sensor 1
470 OUTPUT @Pwr_mtr;CAL1?
480 ENTER @Pwr_mtr;Pass_cal
490 IF Pass_cal=0 THEN PRINT Calibration Passed

Table of Contents

Related product manuals