Power Sensors
Publication 21555, Rev. E, September 2002 B-7
Notes:
1. Includes uncertainty of reference standard and transfer uncertainty. Directly traceable to NIST.
2. Square root of sum of the individual uncertainties squared (RSS).
3. Cal Factor numbers allow for 3% repeatability when connecting attenuator to sensor, and 3% for attenuator measurement
uncertainty and mismatch of sensor/pad combination. Attenuator frequency response is added to the Sensor Cal Factors
which are stored in the sensor’s EEPROM.
For additional specifications, see the Series 80350A (P/N 21568) publication and data sheet.
Table B-4: Peak Power Sensor Cal Factor Uncertainties
Freq. (GHz)
Sum of Uncertainties (%)
1
Probable Uncertainties (%)
2
Lower Upper 80350A
80353A
80354A
80351A
3
80352A
3
80355A
3
80350A
80353A
80354A
80351A
3
80352A
3
80355A
3
0.1 1 1.61 3.06 9.09 9.51 10.16 1.04 1.64 4.92
1 2 1.95 3.51 9.43 9.85 10.50 1.20 1.73 5.04
2 4 2.44 4.42 13.10 13.57 14.52 1.33 1.93 7.09
4 6 2.67 4.74 13.33 13.80 14.75 1.41 2.03 7.17
6 8 2.86 4.94 13.52 13.99 14.94 1.52 2.08 7.25
8 12.4 3.59 6.04 14.25 14.72 15.67 1.92 2.55 7.56
12.4 18 4.09 6.86 19.52 20.97 21.94 2.11 2.83 12.37
18 26.5 —— 9.27 —— —— —— —— 3.63 ——