OPTIC-4 User’s Guide
Issue 5.3
Technical Specifications
Page 13-1
13. Technical Specifications
General
Dimensions: 34 cm 14 cm 34 cm (h w d)
Weight: 8.4 kg (OPTIC-4DC controller)
Ambient temperature range: 18 - 40C
Ambient operating humidity: 40 - 70%
Mains power: 100 - 240 VAC, 50-60 Hz
Typical power consumption: 150 VA
Maximum power consumption: 450 VA
Conforms to safety standards:
The European Low Voltage Directive
The European Electromagnetic Compatibility Directive
RoHS Compliant
Inlet
Full electronic pressure/flow control
Maximum operating temperature: up to 600C at a GC oven temperature of 35C
Minimum operating temperature (without forced cooling): ambient temperature 2C (host
GC oven is OFF)
Cooling: air (down to 35C), LCO
2
(down to -50C), LN
2
(down to -180C)
Temperature ramp rates: 0.1 - 60 C/sec
Up to nine temperature program ramps including negative
Compatible with Merlin MicroSeal septum
Compressed air for cooling: moisture and oil free air at a pressure of 500 kPa.
Electronic Flow Control
Two EFC channels
Full electronic control of column, split and septum purge flows
Pressure range: 7 - 700 kPa
Total flow range: 6 - 500 mL/min He (main channel, all models), 2 - 100 mL/min He (auxiliary
channel, OPTIC-4D and OPTIC-4DC only)
Pressure sensors:
Accuracy: 1% of full scale
Repeatability: 0.2% of full scale
Flow sensors:
Accuracy: total flow - 1% of full scale, septum purge flow - 10% of full scale (1 mL/min)
Repeatability: 0.2% of full scale
Up to nine flow program ramps
He, N
2
or H
2
as carrier gas at a maximum pressure of 700 kPa
Solvent sensor in the split line
Charcoal Filter in the split line