EasyManua.ls Logo

GL Sciences OPTIC-4 - Servicing

GL Sciences OPTIC-4
118 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPTIC-4 User’s Guide
Issue 5.3
Maintenance
Page 11-4
CAUTION!
When leak testing connections, use a suitable electronic leak detector.
Under no circumstances should a soap solution or similar be used as this
will contaminate the inlet!
11. When a leak is detected, tighten the part or remove it and check for a problem. Replace the
part if any problem is found.
CAUTION!
Do not to tighten the inlet base nut excessively. The inlet base is very fragile and
can be damaged easily!
11.4 Servicing
The OPTIC-4 system does not normally require any servicing. If there are problems, which cannot be
solved by replacing consumables, cleaning the inlet, or following the troubleshooting instructions,
then please contact your supplier for further support.
If inlet is temporarily removed from the GC, block its bottom to prevent the inlet interior from
contamination. Use a no-hole graphite ferrule (2406-1019, OPTIC Graphite Ferrule, no-hole, pk/10)
for it.

Table of Contents