GSI TympStar Pro User Manual
D-0129427 Rev. B Page 31 of 162
OPERATING THE TYMPSTAR PRO
PRELIMINARY CHECKS
It is good practice to perform a test on a normal ear each day to make certain that the
instrument is functioning properly. See Routine Maintenance for details. Before starting
any procedures using the GSI TympStar Pro, ensure that the power cord is plugged into a
properly grounded receptacle.
Check also that all cords from the transducers, and printer fit securely in their
connectors on the rear and side panels. Inspect all cords for fraying and
damage. If there is any damage to any cord, do not use the TympStar Pro.
Handle the probe and contra earphones with care. Do not drop them nor allow
them to be banged together. Severe mechanical shock can alter their operating
characteristics or change the output levels, which may require that the
transducers be replaced.
It is recommended that all parts that come into direct contact with the patient
are subjected to standard disinfecting procedures between patients. This
includes physically cleaning and using a recognized disinfectant. Individual
manufacturer’s instructions should be followed for use of any disinfecting agent.
PREPARE TO TEST
Once you have performed the preliminary checks and any routine maintenance checks
prepare to test patients.
√ Turn on the instrument and allow it to come to operating temperature
(approximately 10 minutes).
√ Check that the transducers and other system components are operating properly.
√ Seat the patient comfortably in the test area.
√ Place the selected transducers on the patient.
HELPFUL HINTS
Tympanometry and acoustic reflex testing may be performed on patients of any age;