EasyManuals Logo

GW Instek GDS-2000A Series User Manual

GW Instek GDS-2000A Series
261 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
CONFIGURATION
143
3. Press Attenuation on the side menu
and use the Variable knob to set
the attenuation.
Alternatively, press Set to 10X.
Range
1mX ~1kX (1-2-5 step)
Note
The attenuation factor adds no influence on the real
signal. It just changes the voltage/current scale on the
display.
Set the Deskew
Background
The deskew function is used to compensate for the
propagation delay between the oscilloscope and
the probe.
Panel Operation
1. Press one of the Channel keys.
CH1
2. Press Probe from the bottom menu.
3. Press Deskew on the side menu and
use the Variable knob to set the
deskew time.
Alternatively, press Set to 0s to
reset the deskew time.
Range
-50ns~50ns, 10ps increments
4. Repeat the procedure for another channel if
necessary.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek GDS-2000A Series and is the answer not in the manual?

GW Instek GDS-2000A Series Specifications

General IconGeneral
BrandGW Instek
ModelGDS-2000A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals