MEASUREMENT
81
Delay type measurement is not available in this
mode as only one channel is used as the source.
Use the Individual measurement mode (page 76)
instead.
Only Frequency, Period, +Width, -Width and Duty
Cycle measurements are supported for digital
channels.
High Low Function
The High-Low function is used to select the
method for determining the value of the High-Low
measurement values.
Automatically chooses the best
high-low setting for each
waveform when measuring.
Uses histograms to determine the
high-low values. This mode
ignores any preshoot and
overshoot values. This mode is
particularly useful for pulse-type
waveforms