MAINTENANCE
245
3. Press the CH1 key to activate CH1.
CH1
4. Set the Coupling to DC from the
bottom menu.
5. Set the Probe attenuation to
Voltage, 10X.
Page 142
6. Press the Autoset key. The
compensation signal appears on
the display.
Autoset
7. Press the Display key, then
set the display type to
Vector.
Display
8. Turn the adjustment point on the probe to
make the waveform as square as possible.
Under
Compensation
Normal
Over
Compensation