7.2 Measurement Specifications
1
2
3
4
5
6
7
8
9
10
11
付録
索引
145
Auto range
Preset Test Voltage Measuring Range
Preset test voltage
(setting)
Measuring range
250 V
0.00 M - 250 G
500 V
0.00 M - 500 G
1 kV
0.00 M - 1.00 T
2.5 kV
0.00 M - 2.50 T
5 kV
0.00 M - 5.00 T
Resistance Ranges
When a value below the lower limit of each range is dis-
played, the accuracy is not guaranteed
Resistance range
name
Measuring range
10 M range
0.00 M - 9.99 M
100 M range
9.0 M - 99.9 M
1000 M range
90 M - 999 M
10 G range
0.90 G - 9.99 G
100 G range
9.0 G - 99.9 G
1000 G range
90 G - 999 G
5 T range
0.90 T - 5.00 T

Table of Contents