7.2 Measurement Specifications
146
Temperature and humidity range for guar-
anteed accuracy: 0 to 28°C, less than
90%RH (non-condensation)
Measurement Accuracy
Measuring range
Measurement
accuracy
Equal to or less than the resistance
obtained by dividing the test voltage
(setting) by 100 nA.
± 5%rdg.±5dgt.
Greater than the resistance obtained
by dividing the test voltage (setting)
by 100 nA, and up to 500 G
±20%rdg.±5dgt.
501 G to 5.00 T
±30%rdg.±50dgt.
Test voltage (setting)
Insulation resistance
±30%rdg.±50dgt.
±20%rdg.±5dgt.
±5%rdg.±5dgt.

Table of Contents