EasyManua.ls Logo

Hioki RM3545-02 - Initializing (Reset)

Hioki RM3545-02
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.7 Initializing (Reset)
134
Three reset functions are available.
For more information about communications commands, see the included application disc.
Reset: Returns measurement conditions (except the panel data) to factory defaults.
The instrument can be reset by three methods.
Reset from the System setting screen
Turn on the instrument while holding down and .
Reset by remote control command
RST command (Interface settings are not initialized.)
System reset: Returns all measurement conditions and the panel save data to factory
defaults.
The instrument can be system reset by three methods.
System reset from the System setting screen
Turn on the instrument while holding down , , and .
Reset by remote control command
:SYSTem:RESet command (Interface settings are not initialized.)
Multiplexer channel reset: Returns the multiplexer channel settings to factory
defaults.
The instrument’s multiplexer channels can be reset by two methods.
System reset from the System setting screen
Reset by remote control command
[:SENSe:]CHReset command
This procedure describes reset from the System setting screen.
7.7 Initializing (Reset)
1
Open the Settings Screen.
2
Open the System Setting Screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the [SYS]
tab with the left and right cur-
sor keys.

Table of Contents

Related product manuals