EasyManua.ls Logo

Hioki RM3545-02 - Switching Measurement Currents (100 Mω to 100 Ω)

Hioki RM3545-02
364 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.2 Switching Measurement Currents (100 m to 100 )
66
Power equivalent to the resistance value × (measurement current)
2
will be applied to the
measurement target. If there are any of the following concerns, depending on the level of
the measurement current, set the measurement current to low.
The measurement target may melt (such as a fuse or inflator).
The measurement target may heat up, causing a change in resistance.
The measurement target may be magnetized, causing a change in inductance.
See: "3.1 Checking the Measurement Target" (p.48)
Because the detection voltage decreases when the measurement current is Low, measure-
ment is more susceptible to external noise. If measured values fail to stabilize, take steps to
address the noise, referring to "Appendix 7 Unstable Measured Values" (p. A12). The fol-
lowing four steps are particularly effective in this situation:
Shield the measurement cable (connect the shielding to the instrument’s GUARD termi-
nal).
Twist the measurement cables together.
Shield the measurement target (connect the shielding to the instrument’s GUARD termi-
nal).
Decrease the measurement speed or use the averaging function.
4.2 Switching Measurement Currents
(100 m to 100 )
Range
High Low
Measurement
current
Maximum power
in measurement
range
Measurement
current
Maximum power
in measurement
range
10 m 1 A 12 mW
100 m 1 A 120 mW 100 mA 1.2 mW
1000 m 100 mA 12 mW 10 mA 120 W
10 10 mA 1.2 mW 1 mA 12 W
100 10 mA 12 mW 1 mA 120 W
1000 1 mA 1.2 mW
10 k 1 mA 12 mW
100 k 100 A 1.2 mW
1000 k 10 A 120 W
10 M 1 A 12 W
100 M
(high-
precision mode: ON)
100 nA 1.2 W
100 M, 1000 M
(high-
precision mode: OFF)
1 A or less 1.3 W

Table of Contents

Related product manuals