5.3 Performing Statistical Calculations on Measured Values
112
Turning on the statistical calculation function causes statistics to be calculated based on the
EXT I/O TRIG signal. The timing at which statistics are calculated for measured values
varies with the trigger source setting.
• With external (EXT) triggering: If the TRIG signal is input, one measurement is performed
and subjected to statistical calculation.
• With internal (INT) triggering : If the TRIG signal is input, statistics will be calculated
using the last updated measured value. When using the
auto-hold function, statistics will be calculated using the
held measured value.
Operation is the same in the following cases (except when using auto-hold):
• when pressing the
• when a ∗TRG remote control command is received
When the EXT I/O PRINT signal is input, operation varies with the trigger source.
• When using an external trigger [EXT]: The most recent measurement results are printed.
• When using the internal trigger [INT]: Statistics are calculated using the last updated
measured value and printed after the PRINT signal is input.
• The same operation can be accomplished by pressing [PRINT] on the MENU P.1/3
display.
Using Statistical Calculations
1
Open the Settings Screen.
2
Open the System Setting Screen.
The Settings screen
appears.
Switch the function menu
to P.2/3.
1
2
Move the cursor to the [SYS]
tab with the left and right cur-
sor keys.